Data for Publication: ChemPlusChem
sample designation in the publication: M; M-N1; M-N2; M-N3; M-N4; M-N5
File:ASAP
File format: *xlsx;*csv
Figure 1, Table 1 in manuscript
Data from nitrogen adsorption/desorption measurments.
ASAP2020 Micromeritics; ASAP2020 softwere
Samples degassed at 200 ◦C under vacuum for 24 h. Nitrogen adsorption at −196◦C. The specific surface area calculated using Braunauer, Emmet and Teller (BET) equation; micropore size using the Dubinin–Astakhov equation.
File:FTIR
File format: *csv
Figure 3 in manuscript
Data from FTIR measurments
NICOLET 6700 spectrometer; Thermo Fisher Scientific; OMNIC softwere; DRIFT (5% of sample in KBr);
Spectra range 400–4000 cm−1; resolution 4 cm−1.
File:TG
File format: *csv
Figure 4, Table 2 in manuscript
Data from thermogravimetric measurments
Star 850, Mettler Toledo. Temperature range of 25-800 °C, heating ramp of 10 deg/min, air rate 60 ml/min
File:Catalytic Experiments
File format:*pdf
Figure 5-10, Table 4 in manuscript
Reports from GC chromatography
Agilent 7890A, flame ionization detector FID; capillary column HP5; Inlet temperature: 250◦C ; sample volume: 1um, split ratio: 5:1; oven temp: 40◦C-200o◦C He flow 6.5 ml/min; detector: temperature 300◦C; H2 flow:30 ml/min, Air flow:400ml/min