In order to establish the conformity of the ADS1298ECGFE-PDK device with the ECG standard PN-EN 60601-2-25 p. 201, the following measurements were carried out:
1.Input impedance (12.4.103) (should be greater than 2.5 MΩ) - files 13111-13119 and 13121-13129.
2.ECG path sensitivity (12.4.104) (should be changed stepwise with an accuracy of 5% in the presence of differential and common voltage of ±300 mV) - files 14111-14119.
3.Resistance to overdrive (12.4.105.2)(differential voltage of 1 V for 10 s) - files 41111-41119 and 42111-42119.
4.Noise level (12.4.106.1)(at with only the mains filter switched on, should not exceed 30 uV) - files 91110-91119.
5.Channel crosstalk (12.4.106.2)(should not exceed an amplitude of 0.5 mm at normal sensitivity, i.e., 2% of the amplitude of the given signal multiplied by the set sensitivity) - files 61111-61119.
6. Frequency response (12.4.107.1)(in the frequency range up to 150 Hz) - high frequency (A): files 15111-15119, 15121-15129 and 24111-24119; high frequency (E): files 21111-21119 and 21211-21219; low frequency: files 22111-22119.
7. Linearity and range of transfer (12.4.107.2)(should be ensured in each lead for 20 mV with an accuracy of 5% over the entire width of the recording field, also in the presence of a common or differential voltage of 300 mV of any polarity) - files 101111-101119.